beautypg.com

13 update logic statement, 14 ic6 event list – INFICON IC6 Thin Film Deposition Controller User Manual

Page 191

background image

10 - 21

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.13 Update Logic Statement

Command = UL elements>

Statement Number = 1 - 100

Length of logic elements = number of bytes of the Logic element for each
statement, including the 0x03 (ETX) terminator.

Set of elements = See description in QL above.

Example: IF External Input 21 and External Input 2 and External Input 18 and
External Input 8

THEN Start Timer 1 and Extern Out On 10

Hex code for above including length and checksum:
1500554C01114115264102264112264108205101410A031A

Response = None (Just header and trailer)

Logic changes are not allowed while a process is running or if the IC6 is in a stop.

10.4.14 IC6 Event List

Events used in the Logic Statement IF section. Logic codes are used in the
remote communication query and update commands. The Logic codes start at
0x41 because there are formatting codes (for connectors, parentheses, etc.)
used in the logic statements that are ASCII codes below 0x41. See

Table 10-7

.

Table 10-7 Event list

Events

1st

numeric

(1 Byte)

(XXX)

Min Max

2nd

numeric

(Integer)

(YYY)

Min

Max

Logic

Code

Logic

Code

Negated

Parenthesis

External Input

Input #

1

28

0x41

0xBF

Pre-Deposit

0x42

0xBE

In Deposit

0x43

0xBD

Post-Deposit

0x44

0xBC

Non Deposit

0x45

0xBB

Ready

0x46

0xBA

Crucible Switch

0x47

0xB9

PreCond Rise

0x48

0xB8

PreCond Soak

0x49

0xB7

Rise 1

0x4A

0xB6