beautypg.com

18 query material name, 19 update material name, 20 query output name – INFICON IC6 Thin Film Deposition Controller User Manual

Page 205: 21 update output name

background image

10 - 35

PN

07

4-

50

5-

P1

F

IC6 Operating Manual

10.4.18 Query Material Name

Command = QN

Material Number = 1-32

Response = < Material Name>

Material Name = Up to 15 characters, null terminated. The string returned
will be a blank if the default Material # name has not been changed.

10.4.19 Update Material Name

Command = UN < Material Number> < Material Name>

Material Number = 1 - 32

Material Name = Up to 15 characters, null terminated

Response = None (Just header and trailer)

10.4.20 Query Output Name

Command = QO

Output Number = 1 - 38

Response =

Output Name = Up to 15 characters, null terminated

10.4.21 Update Output Name

Command = UO

Output Number = 1 - 38

Output Name = Up to 15 characters, null terminated

Response = None (Just header and trailer)

Output Names may only be updated if the IC6 is in Ready or Stop.

NOTE: UO changes are not allowed while a process is running. The UO command

will not allow updating a name set by hardware designation, unless it is
identical to the defined hardware name