IC6 Operating Manual

10

(0x0a)

Deposit Average

Rate

4 or 24

KÅ Based on Averaging Time

parameter. Only available when in a

control state.

11

(0x0b)

Deposit Average

Power

4 or 24

% Based on Averaging Time

parameter. Only available when in a

control state.

12

(0x0c)

Raw Rate

4 or 24

Å/S

13

(0x0d)

One Minute

Averages for Rate &

Power

8 or 48

Returns a 'No Data Error' if not enough
data to do a one minute average of the

rate or the power

14

(0x0e)

Rate Deviation /

power output trend

Varies with

# of data

points

Number of bytes in the response.

Average values from the operate screen

graph.

Format of each data point.

15

(0x0f)

State before Suspend

1 or 6

See encodes in

section 3.5 on page

3-32

. If not currently in Suspend, will

return "No Data Error"

16

(0x10)

Data Log

76 or 456

Multiple pieces of

information

See

section 3.6.10 on page 3-43

for

order of data.

17

(0x11)

Data Log with Xtal

History

3,148

or

18,888

for 12 crystals in

8 sensors>.

See

section 3.6.10 on page 3-43

for

order of data. Length of response = 76

bytes Data Log +

3072 bytes Xtal History

Xtal History = 32 bytes per crystal, 12
crystals in 8 sensors = 33*12*8=3072

bytes

Table 10-13 Status layer (continued)

SL

Command

ID

Description

Number of

bytes in

the

response

Data Returned

Notes