33 remote general action, Mmands, see – INFICON IC6 Thin Film Deposition Controller User Manual
Page 217

10 - 47
PN
07
4-
50
5-
P1
F
IC6 Operating Manual
10.4.33 Remote General Action
Command = RG
Command ID =
.
Action Value =
See
for commands that require an action value.
Response = None (Just header and trailer)
Table 10-15 Remote general
RG
Command
ID
Meaning
Action Value
Notes
0 (0x00)
Start
Duplicates the equivalent front panel action
1 (0x01)
Stop
Duplicates the equivalent front panel action
2 (0x02)
Reset
Duplicates the equivalent front panel action
3 (0x03)
Remote Lock On
No parameters can be updated from the front panel
"R Lock" displayed
4 (0x04)
Remote Lock Off
Cancel RG 3 command
5
(0x05)
Set Logic Statement
to True
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
6
(0x06)
Clear Logic
Statement
to False.
Range is 1-100.
Only does something if logic statement event "Computer
Control" set. Otherwise, ignored.
7
(0x07)
Crystal Fail Inhibit On
Range is 0-8 (0 = All)
"Crystal Fail" output always set to FALSE, even during crystal
fail
8
(0x08)
Crystal Fail Inhibit Off
Range is 0-8 (0 = All)
"Crystal Fail" output has normal meaning
9
(0x09)
Set date (Month, Day,
Year) or (Day, Month,
Year) Depending on
General parameter
17 (Date Format).
Range as follows:Month = 1 to 12
Day = 1 to 31
Year = 2000 to 2099
10
(0x0a)
Set time (Hour,
Minute, Second)
Range as follows Hour = 0 to 23
Minute = 0 to 59
Second = 0 to 59
11
(0x0b)
Zero Counter
Counter Range is 0-20
0 = All
12
(0x0c)
Zero Timer
Timer Range is 0-20
0 = All