11 Device manager

900 Touch Control

■■■■■■■■

79

11 Device manager

Main dialog: System

Device manager

This chapter describes how you can configure the Touch Control, the con-
trol instruments (Titrando, USB Sample Processor etc.) and the connected
peripheral devices. A detailed description of the hardware installation can
be found in the installation manual for the control instrument.

The instrument list gives the name and type of every instrument. Connec-
ted control instruments (Titrando, Dosing Interface, USB Sample Pro-
cessor, etc.) with the peripheral devices (dosing devices, stirrers, etc.) con-
nected to the MSB connector are entered in the list automatically. A USB/
RS-232 adapter will also be automatically recognized and entered in the
instrument list with default settings. A printer is entered in the instrument
list in the default settings. You must enter a PC keyboard, a barcode
reader or a balance in the instrument list yourself.

[New]

Add a new device to the list (see Chapter 11.1, page 80).

[Delete]

Delete the selected device from the list.

NOTE

Devices which are recognized automatically cannot be deleted from the
list while they are still connected.

[Edit]

Configure the selected device (see Chapter 11.2, page 80).