29.4 Volumetric Karl Fischer titrations (KFT)
■■■■■■■■■■■■■■■■■■■■■■
322
■■■■■■■■
900 Touch Control
Start volume
Volume that is dosed prior to the start of the titration.
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range
0 - 999999 s
Default value
0 s
Initial measured value
The conditions for the entry of the first measured value in the measuring
point list are defined under [Initial meas. value].
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.
Measuring mode Ipol:
Input range
0.1 - 999.0 mV/min
Selection
off
Default value
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Upol:
Input range
0.01 - 99.90 µA/min
Selection
off
Default value
off