beautypg.com

2 dosing fixed volumes, Dosing fixed volumes – Metrohm 900 Touch Control Manual User Manual

Page 274

background image

28.3 Dosing

■■■■■■■■■■■■■■■■■■■■■■

258

■■■■■■■■

900 Touch Control

dynamic
This setting can only be selected when the dosing cylinder is filled. The
dosing is being carried out faster and faster until the maximum dosing
rate is reached (starting with 1 mL/min and then doubling the dosing
rate every 1.5 s).

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488)
.

28.3.2

Dosing fixed volumes

Manual control: Dosing

Dosing fixed vol.

You can dose a particular volume with the [Dosing fixed vol.] function.

Proceed as follows:

1 Configure the dosing function

NOTE

The dosing and filling rates should be decreased for viscous liq-
uids.

The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 32, page 488).

Enter the desired volume.