background image

29.4 Volumetric Karl Fischer titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

328

■■■■■■■■

900 Touch Control

Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Stop time

The titration is stopped when the specified time has elapsed following the
termination of the start conditions.

Input range

1 - 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488)
.

29.4.5

Conditioning

The conditions required for conditioning are defined under [Condition-
ing]
.

Conditioning

on | off (Default value: on)

If this parameter is activated, then the first time the titration is started the
working medium will be titrated to the endpoint with the specified control
parameters. The status is kept stable. The actual method run does not
begin until [

] has been pressed once more.

Start drift

Conditioning OK will be displayed as soon as this drift has been reached
and the titration can be started.

Input range

1 - 999 µL/min

Default value

20 µL/min

Drift correction

The endpoint volume can be corrected by taking the drift value into
account. For this, the drift is multiplied with the drift correction time and
this value is then subtracted from the endpoint volume. The drift correc-
tion time is the time interval between the end of conditioning and the end
of the determination.