29 Parameters

900 Touch Control

■■■■■■■■

357

-
Negative measured value modification, i.e. in the direction of a lower
pH value or lesser voltage.
auto
The titration direction is determined automatically from the start mea-
sured value and the set control point.

Temperature

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command). This value is used for temperature correction in pH
measurements.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Time interval MP

Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.

Input range

0.1 - 999999.0 s

Default value

2.0 s

Start time

The measured values are not entered in the measuring point list until this
time has elapsed since the start of the titration.

Input range

0 - 999999 s

Default value

0 s

Start meas. value

The measured values are not entered in the measuring point list until this
start measured value has been reached.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U:
Input range

–2,000.0 - 2,000.0 mV