29 Parameters

900 Touch Control

■■■■■■■■

341

29.5.6

Control device

The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System

Device

manager.

NOTE

This button is only displayed when several control devices have been
configured.

Control device

Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.

Selection

Selection of configured control devices

29.5.7

Sensor

The parameters for the sensor are edited under [Sensor].

Sensor

Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System

Sensors. You can

also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.

Selection

Selection of configured sensors

I(pol)

The polarization current is the current that is applied to a polarizable elec-
trode during the voltametric measurement.

Selection

5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

Electrode check

on | off (Default value: off)

For polarizable electrodes and the generator electrode, an electrode check
can be carried out. The following will be tested: