10 Sensors

900 Touch Control

■■■■■■■■

77

If this parameter is activated, then the cell constant will be monitored.

Lower limit

Input range

0.001 - 500.000 /cm

Default value

0.850 /cm

Upper limit

Input range

0.001 - 500.000 /cm

Default value

1.150 /cm

10.6

Monitoring the calibration interval (only for pH and
ion-selective electrodes and conductivity measuring
cells)

Sensor: Edit

Calibration interval

In the dialog Edit sensor / Calibration interval, you can define the time
interval after which the sensor must be recalibrated.

Calibration date

Date of the last calibration.

Monitoring

on | off (Default value: off)

If this parameter is activated, then the validity of the calibration will be
monitored.

Calibration interval

If you define a time interval for the validity of the calibration, then the
date in Next calibration will be tracked automatically.

Input range

1 - 999 days

Default value

7 days

Next calibration

If you define a date for the next calibration, then the Calibration inter-
val
will be tracked automatically.

Format: YYYY:MM:DD

Action

Selection of the action which is carried out when the time interval has
expired.