29.7 STAT titrations (STAT)

■■■■■■■■■■■■■■■■■■■■■■

358

■■■■■■■■

900 Touch Control

Selection

off

Default value

off

Start rate

The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. Checking this parameter is not
started until 10 s after the start of the command.

Input range

0.01 - 166.00 mL/min

Selection

off

Default value

off

NOTE

If several of the parameters Start time, Start meas. value or Start
rate
have been defined, then all of them must be fulfilled for measured
values to be entered in the measuring point list.

29.7.4

Stop conditions

The conditions for canceling the titration are defined under [Stop condi-
tions]
.

Stop volume

The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.

Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Stop time

The titration is stopped when the specified time has elapsed.

Input range

0 - 999999 s

Selection

off

Default value

off

Since

This parameter can only be edited when a stop time has been defined.

This parameter defines the start of the stop time.

Selection

Start | Control point | Last dosing