29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
358
■■■■■■■■
900 Touch Control
Selection
off
Default value
off
Start rate
The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. Checking this parameter is not
started until 10 s after the start of the command.
Input range
0.01 - 166.00 mL/min
Selection
off
Default value
off
NOTE
If several of the parameters Start time, Start meas. value or Start
rate have been defined, then all of them must be fulfilled for measured
values to be entered in the measuring point list.
29.7.4
Stop conditions
The conditions for canceling the titration are defined under [Stop condi-
tions].
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range
0.00000 - 9999.99 mL
Default value
100,000 mL
Selection
off
Stop time
The titration is stopped when the specified time has elapsed.
Input range
0 - 999999 s
Selection
off
Default value
off
Since
This parameter can only be edited when a stop time has been defined.
This parameter defines the start of the stop time.
Selection
Start | Control point | Last dosing