28.3 Dosing

■■■■■■■■■■■■■■■■■■■■■■

262

■■■■■■■■

900 Touch Control

28.3.5

Filling

Manual control: Dosing

Fill

You can use the function [Fill] to fill the dosing cylinder of the exchange
unit/dosing unit manually. The filling rate cannot be configured; the maxi-
mum filling rate will be used.

28.3.6

Replacing reagent

NOTE

You can easily change the reagent in a dosing unit without no contact
with the chemicals using the functions Prepare and Empty.

Proceed as follows:

1 Execute the function [Empty].

2 Attach the dosing unit to the bottle with a suitable solvent.

3 Execute the function [Prepare].

The last traces of the reagent in the cylinder and tubings are
removed with the solvent.

4 Execute the function [Empty] again.

5 Attach the dosing unit to the bottle with the new reagent.

6 Execute the function [Prepare].

The dosing unit can now be used.