Metrohm 900 Touch Control Manual User Manual
Page 283
■■■■■■■■■■■■■■■■■■■■■■
28 Manual control
900 Touch Control
■■■■■■■■
267
Tower for the manual control.
Selection
Tower 1 | Tower 2
Tower 1
View from the front, the right tower.
Tower 2
View from the front, the left tower. This selection is only available for
Sample Processors with two towers.
Lift position
Current lift position.
Rack position
Current rack position. If the rack is not set to a valid position, e.g. after a
rack reset, -- is displayed to indicate this.
NOTE
Note that the rack has to be set to a valid position for all lift move-
ments! Otherwise a corresponding error message will be displayed.
Move to a valid rack position with [Previous] or [Next]. Alternatively,
you can carry out a MOVE command.
[Home position]
Move the lift to the home position (corresponds to lift position 0 mm).
[Work position]
Move the lift to the work position of the current rack position or the
external position.
The work position can directly be defined here (see "Assigning lift posi-
tions", page 269) or in the device manager (see "Lift positions Tower
1/2", page 103).
[
⇧]
Continuously move the lift upwards as long as the button is pressed
down. The lift rate used is defined in the dialog Sample Processor / Lift.
[
⇩]
Continuously move the lift downwards as long as the button is pressed
down. The lift rate used is defined in the dialog Sample Processor / Lift.
[Previous]
Move to the previous rack position.
[Next]
Move to the next rack position.