29 Parameters

900 Touch Control

■■■■■■■■

349

NOTE

This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working
medium.

29.6.5

Cell

The parameters for the titration cell are defined under [Cell].

Generator electrode

Type of generator electrode.

Selection

without diaphragm | with diaphragm

Default value

with diaphragm

without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection

100 mA | 200 mA | 400 mA | auto

Default value

auto

auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint.

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System

Reagents.

Selection

Selection of configured reagents | off

Default value

off

off
The reagent monitoring is not active.