29 Parameters

900 Touch Control

■■■■■■■■

451

29.15.5.2

Stop conditions

The conditions for canceling the dosing are defined under [Stop condi-
tions]
.

Stop volume

The dosing is stopped when the specified volume has been dosed since
the start of the dosing. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.

Input range

0.00000 - 99999.9 mL

Selection

off

Default value

off

Stop time

The dosing is stopped when the specified time has elapsed since the start
of the dosing.

Input range

0 - 999999 s

Selection

off

Default value

off

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488)
.