29 Parameters

900 Touch Control

■■■■■■■■

461

Use the highest possible filling rate in order to keep the filling times as
short as possible. The filling rate should be decreased for viscous liq-
uids.

When you use two buret units with different dosing cylinders, the fill-
ing rate for the larger sized cylinder must be at the minimum:

= filling rate in mL/min for the larger sized cylinder

= filling rate in mL/min for the smaller cylinder

= cylinder volume in mL of the buret unit of the second dosing
device

= cylinder volume in mL of the buret unit of the first dosing
device

Example:

Dosing device 1: volume = 20 mL, filling rate = 50 mL/min

Dosing device 2: volume = 50 mL

Filling rate 2

≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min

The dosing rate must not exceed 75 % of the filling rate of the smaller
cylinder. These values are listed in the following table, valid at maxi-
mum filling rate:

Table 19

Maximum dosing rate for different dosing cylinders

Cylinder volume

maximum dosing rate

Exchange unit

Dosing unit

1 mL

2.25 mL/min

2 mL

5.00 mL/min

5 mL

11.25 mL/min

12.50 mL/min

10 mL

22.50 mL/min

25.00 mL/min

20 mL

45.00 mL/min

50.00 mL/min

50 mL

112.50 mL/min

124.50 mL/min