Use the highest possible filling rate in order to keep the filling times as
short as possible. The filling rate should be decreased for viscous liq-
uids.
■
When you use two buret units with different dosing cylinders, the fill-
ing rate for the larger sized cylinder must be at the minimum:
= filling rate in mL/min for the larger sized cylinder
= filling rate in mL/min for the smaller cylinder
= cylinder volume in mL of the buret unit of the second dosing
device
= cylinder volume in mL of the buret unit of the first dosing
device
Example:
Dosing device 1: volume = 20 mL, filling rate = 50 mL/min
Dosing device 2: volume = 50 mL
Filling rate 2
≥ 50 mL/min · 50 mL / 20 mL ≥ 125 mL/min
■
The dosing rate must not exceed 75 % of the filling rate of the smaller
cylinder. These values are listed in the following table, valid at maxi-
mum filling rate:
Table 19
Maximum dosing rate for different dosing cylinders
Cylinder volume
maximum dosing rate
Exchange unit
Dosing unit
1 mL
2.25 mL/min
–
2 mL
–
5.00 mL/min
5 mL
11.25 mL/min
12.50 mL/min
10 mL
22.50 mL/min
25.00 mL/min
20 mL
45.00 mL/min
50.00 mL/min
50 mL
112.50 mL/min
124.50 mL/min