28 Manual control

900 Touch Control

■■■■■■■■

253

Dialog "Measure / Temperature compensation"
The parameters for converting the measured conductivity to a reference
temperature are defined under [Temp. compens.].

Reference temp.

The measured conductivity is converted to this temperature.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Selection

off

off
The temperature compensation is disabled.

Temp. compensation

Type of temperature compensation.

Selection

constant | ISO / DIN

constant
The temperature compensation is carried out with the value entered
for the temperature coefficient.
ISO / DIN
For ground water, spring water and surface water, the temperature
coefficients according to standard ISO 7888:1985 (German version:
DIN EN 27888:1993) are stored in the system.

Temp. coefficient

This parameter can only be edited with Temp. compensation = con-
stant
.

Temperature coefficient belonging to the reference temperature entered
above. With this temperature coefficient, the conductivity is converted to
the reference temperature.

Input range

0.00 - 9.99 %/°C

Default value

0.00 %/°C