29.9 Measurements (MEAS)

■■■■■■■■■■■■■■■■■■■■■■

384

■■■■■■■■

900 Touch Control

ted data chip, the cylinder volume and the type of dosing drive are
checked. For the selected standard solution, the validity of the titer, the
working life of the standard solution and the time interval for the GLP test
of the buret unit are checked at the start of the determination.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place.

Dosing rate

Three predefined sets of parameters can be selected for the dosing rate.
The maximum dosing rate depends on the cylinder volume (see Chapter
32, page 488)
.

Selection

slow | medium | fast

Default value

fast

slow
The dosing rate is at 10 % of the maximum dosing rate.
medium
The dosing rate is at 50 % of the maximum dosing rate.
fast
The dosing rate is equivalent to the maximum dosing rate.

Delta U

This parameter can only be edited with Addition = auto.

The individual volume steps are calculated in the way that the voltage dif-
ference entered is reached.

Input range

1 - 999 mV

Default value

10 mV

Addition increments
The volume of the individual volume steps is defined under [Addition
increments]
. This button is enabled only with Addition = auto dos or
manual.

Volume

Volume of a single volume step.

Input range

00.000 - 9999.900 mL

Default value

00.100 mL