29.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
310
■■■■■■■■
900 Touch Control
Measuring mode pH, U and Ipol:
Input range
0.1 - 999.0 mV/min
Selection
off
Default value
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Upol:
Input range
0.01 - 99.90 µA/min
Selection
off
Default value
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range
0 - 999999 s
Default value
1 s
29.3.2
Control parameters
The control parameters for endpoint 1 and endpoint 2 are defined under
[Control parameters].
Endpoint 1 at
Measured value for the first endpoint.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.