11.5 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

900 Touch Control

Properties which apply only to the selected external position can be
defined in the dialog Properties / Edit external position.

Angle

Swing angle for the selected external position.

Input range

(Offset) - (Offset + max. swing range) °

Default value

60.00 °
The offset is made up of a design-dependent angle
(approx. 8…9°) together with the robotic arm off-
set from the robotic arm properties. The maximum
swing range is also defined under the robotic arm
properties (see "Properties – Robotic arm", page
96)
.

Work position

Work position for the selected external position.

Input range

0 - 'maximum stroke path' mm
A lift position of 0 mm corresponds to the "home
position", i.e. the lift is located at the upper stop
position. The maximum stroke path is defined in the
properties of the tower (see "Maximum stroke
path", page 93)
.