16.5 Method options

■■■■■■■■■■■■■■■■■■■■■■

174

■■■■■■■■

900 Touch Control

NOTE

Do not fail to deactivate this option for the currently loaded method if
you are conditioning a titration cell at the same time in manual control.
Otherwise, the stirrer of the titration cell will also be switched off if the
currently loaded method is canceled. This would result in an overtitra-
tion of the working medium.

Set remote lines

Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System

Templates Output

lines.

NOTE

A line set active is not being reset automatically, not even at the end of
the determination.

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.
If you wish to set pulses with other lengths, you have to define a corre-
sponding template.

Entry

Bit pattern containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

**************

Selection

Selection of the templates defined

16.5.3

Sample data

Main dialog: Edit parameters

Method options Sample data

You can define the following method-specific settings in the dialog
Method options / Sample data:

Modify the designation for the sample identifications

Define a fixed sample size