3 sample data, Sample data – Metrohm 900 Touch Control Manual User Manual
Page 190
16.5 Method options
■■■■■■■■■■■■■■■■■■■■■■
174
■■■■■■■■
900 Touch Control
NOTE
Do not fail to deactivate this option for the currently loaded method if
you are conditioning a titration cell at the same time in manual control.
Otherwise, the stirrer of the titration cell will also be switched off if the
currently loaded method is canceled. This would result in an overtitra-
tion of the working medium.
Set remote lines
Selection of the signal out of the templates or entering the required bit
pattern. Templates are defined under System
▶ Templates ▶ Output
lines.
NOTE
A line set active is not being reset automatically, not even at the end of
the determination.
Entering the bit pattern:
■
0 = line inactive
■
1 = line active
■
* = retain line status
■
p = set pulse
The output lines are always numbered from right to left, i.e. with the sig-
nal *************1 line 0 is set. With a pulse, the length is set to 200 ms.
If you wish to set pulses with other lengths, you have to define a corre-
sponding template.
Entry
Bit pattern containing exactly 14 characters
or a max. of 24 characters for the name of
the template
Default value
**************
Selection
Selection of the templates defined
16.5.3
Sample data
Main dialog: Edit parameters
▶ Method options ▶ Sample data
You can define the following method-specific settings in the dialog
Method options / Sample data:
■
Modify the designation for the sample identifications
■
Define a fixed sample size