11.5 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
92
■■■■■■■■
900 Touch Control
[Racks]
Configure sample racks (see Chapter 11.6, page 99).
[Running time]
Configure the running time meter.
The running time meter adds together the time while the USB Sample Pro-
cessor is "busy" i.e. when a single action is being carried out. A message
with the prompt to service the device is displayed after expiry of the time
limit set here. We recommend to carry out a service after 1000 operating
hours. A reset of the running time meter can only be carried out by a
Metrohm service technician.
[Properties]
Open the properties dialog of the highlighted entry.
11.5.1
Properties – Sample Processor
List of devices: Sample Processor
▶ Edit ▶ Properties
Device name
This designation is used for identification purposes when selecting control
devices (command, manual control).
Entry
24 characters maximum
Comment
Entry
24 characters maximum
Program version
Program version of the instrument software.
Serial number
Shows the serial number of the device.
Rack name
Name of the currently attached sample rack.
[Adjustment data]
Display the internal adjustment data (EEPROM data) of the USB Sample
Processor.
CAUTION
Do not modify and save these settings under any circumstances. This
data is required by the service technician in case of positioning prob-
lems.