beautypg.com

5 note, 6 saving a determination automatically, Note – Metrohm 900 Touch Control Manual User Manual

Page 196: Saving a determination automatically

background image

16.5 Method options

■■■■■■■■■■■■■■■■■■■■■■

180

■■■■■■■■

900 Touch Control

[Sign]

Sign the method or determination. This button is active only when work-
ing with activated login function and password (see Chapter 17.1, page
182)
.

16.5.5

Note

Main dialog: Edit parameters

Method options Note

In this dialog, you can enter a short text, e.g. important information about
carrying out the determination.

[Display options]

Definition when the note is being displayed.

Automatically after loading the method

on | off (Default value: off)

If this parameter is activated, the note is displayed when loading the
method. It can otherwise only be read in this dialog.

16.5.6

Saving a determination automatically

Main dialog: Edit parameters

Method options Save automat.

In the dialog Method options / Save automatically, you can specify
whether the determination is to be saved automatically and/or whether a
PC/LIMS report is to be created.

Save determination automatically

on | off (Default value: off)

If this parameter is activated, then the determination will be saved auto-
matically.

Memory

Memory location where the determination will be saved.

Selection

External memory 1 | External memory 2 |
Shared memory

Default value

External memory 1

Shared memory
The determination will be saved in a shared directory on the network.
The shared directory is selected in the Edit device / Shared memory
dialog (see Chapter 11.3.3, page 84).

Group

Directory where the determination will be saved.

Entry

32 characters maximum