Metrohm 900 Touch Control Manual User Manual
Page 394
29.9 Measurements (MEAS)
■■■■■■■■■■■■■■■■■■■■■■
378
■■■■■■■■
900 Touch Control
Measuring mode T:
Input range
0.1 - 999.0 °C/min
Default value
0.5 °C/min
Selection
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Measuring mode Cond:
Input range
0.001 - 999.000 (mS/cm)/min
Selection
off
Default value
off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed.
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.
Input range
0 - 999999 s
Default value
0 s
Max. waiting time
If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed. As default value, a suitable waiting time for the signal drift is
automatically calculated.
Input range
0 - 999999 s
The default value depends on the measuring mode.
Selection
off
off
The measurement is continued endlessly.
Stop meas. value
The measurement is canceled when the specified measured value has
been reached since the start of the measurement.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.