29 Parameters

900 Touch Control

■■■■■■■■

361

Measuring mode U:
Input range

0.0 - 2,000.0 mV

Default value

2.0 mV

Dosing rate:
Input range

0.00 - 166.00 mL/min

Default value

0.20 mL/min

Temperature:
Input range

0.0 - 150.0 °C

Default value

0.2 °C

Upper limit

When this value is exceeded, the action defined in the following is trig-
gered.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

–20.000 - 20.000

Default value

20.000

Measuring mode U:
Input range

–2,000.0 - 2,000.0 mV

Default value

2,000.0 mV

Dosing rate:
Input range

0.00 - 166.00 mL/min

Default value

166.00 mL/min

Temperature:
Input range

–20.0 - 150.0 °C

Default value

150.0 °C

Upper hysteresis

The upper hysteresis represents the tolerance range for the upper limit
(see Figure 18, page 364). When Action = Wait, the dosing does not
continue until the monitored quantity has once again fallen below the
upper limit incl. the value entered here.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.