1 fixed endpoint evaluation (eval fix-ep) – Metrohm 900 Touch Control Manual User Manual

Page 423

background image

■■■■■■■■■■■■■■■■■■■■■■

29 Parameters

900 Touch Control

■■■■■■■■

407

EVAL FIX-EP (fixed endpoints)
Measured quantities:

– Measured value
– Time
– Volume

EVAL MIN/MAX (minimum/maximum)

EVAL RATE (average dosing rate)

Additional evaluations for measurements
The following additional evaluations are possible:

EVAL FIX-EP (fixed endpoints)
Measured quantities:

– Measured value
– Time

EVAL MIN/MAX (minimum/maximum)

EVAL BREAK (break points)

NOTE

The commands EVAL are not available for concentration measurements
(MEAS Conc).

29.12.1 Fixed endpoint evaluation (EVAL FIX-EP)

For the fixed endpoint evaluation, the associated values are interpolated
from the measuring point list for one quantity (measured value, volume,
etc.). Up to nine fixed endpoints can be evaluated with one command.

Fixed quantity

Selection of the quantity to which the associated value is interpolated
from the measuring point list.

Selection

Measured value | Time | Volume | Quantity

Default value

Measured value

Quantity
Amount of iodine/bromine created (only for KFC/BRC).

Fixed EP1 at

The fixed endpoint must lie between the first and the final entry in the
measuring point list.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.