3 endpoint titrations (set), 1 start conditions, 3 endpoint titrations (set) 29.3.1 – Metrohm 900 Touch Control Manual User Manual
Page 325: Start conditions
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
309
29.3
Endpoint titrations (SET)
29.3.1
Start conditions
The parameters that are carried out before the start of titration are
defined under [Start conditions].
Pause 1
Waiting time, e.g. for the electrode to settle down before a start volume is
added.
Input range
0 - 999999 s
Default value
0 s
Start volume
Volume that is dosed prior to the start of the titration.
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range
0 - 999999 s
Default value
0 s
Initial measured value
The conditions for the entry of the first measured value in the measuring
point list are defined under [Initial meas. value].
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.