29 Parameters
900 Touch Control
■■■■■■■■
385
Apply this value for all increments
on | off (Default value: off)
If you activate this parameter, the entered volume is applied for all volume
steps.
29.9.3
Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button is only displayed when several control devices have been
configured.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-
ces do not have to be connected. This allows to transfer methods more
easily from one system to another.
Selection
Selection of configured control devices
29.9.4
Sensor
The parameters for the sensor are edited under [Sensor].
The parameters are valid for all measuring modes except for Temp. mea-
surement. This parameter is not available with the MEAS T command
(temperature measurement).
Measuring input
Selection of the measuring input the sensor is connected to. The selection
is not dependent on whether the control device has one or two measur-
ing interfaces.
Selection
1 | 2
Default value
1
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System
▶ Sensors. You can
also enter a sensor name which is not contained in the sensor list. When a
determination is started there is a check whether the sensor is contained
in the sensor list.