2 control device – Metrohm 900 Touch Control Manual User Manual
Page 417
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
401
Remote
For automatic calibrations with Sample Processors, which are connec-
ted via the Remote Box.
USB
For automatic calibrations with a USB Sample Processor (see Chapter
32.6, page 492).
Electrode type
Selection of the electrode type.
Selection
Standard | Gel | Non-aqueous | Custom
Default value
Standard
Standard
Electrodes which contain e.g. KCl solution as reference electrolyte.
Gel
Electrodes which contain Idrolyte as reference electrolyte.
Non-aqueous
Electrodes, which contain a non-aqueous reference electrolyte, e.g.
TEABr in ethylene glycol or LiCl in ethanol.
Custom
Under System
▶ Templates a custom electrode type can be defined.
Temperature
Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement = automatic or continuous is defined (see sensor dialog
of the command).
Input range
–20.0 - 150.0 °C
Default value
25.0 °C
29.11.2 Control device
The control device the determination is carried out with is selected under
[Control device]. Control devices are defined under System
▶ Device
manager.
NOTE
This button is only displayed when several control devices have been
configured.
Control device
Selection of the control device from the list of devices. Only those devices
are displayed which are able to carry out the command. The control devi-