4 password options, Password options – Metrohm 900 Touch Control Manual User Manual
Page 51
■■■■■■■■■■■■■■■■■■■■■■
7 System settings
900 Touch Control
■■■■■■■■
35
Password required
on | off (Default value: off)
If this option is activated, then the user must enter a password in addition
to his or her user name or identification profile.
Logout automatically
on | off (Default value: off)
If this option is activated, then the user will be logged out automatically
after the specified time.
Input range
1 - 60 min
Login only for the same user
on | off (Default value: off)
If this option is activated, then only the same user may log in again after
he or she has logged out. Users with administrator rights can, however,
log in at any time.
[Audit Trail]
Define the settings for recording an Audit Trail (see Chapter 7.3.7, page
39).
[Reasons]
Predefine a list of reasons that can be selected at the modification/signa-
ture of a method or determination (see Chapter 7.3.6, page 38).
[Modific. options]
Define for which modifications a reason is required (see Chapter 7.3.5,
page 37).
[Password options]
Define the settings for the password, see following chapter.
7.3.4
Password options
Main dialog: System
▶ System settings ▶ User admin. ▶ Login
options
▶ Password options
You can make various settings for password entry in the password
options.