29 Parameters

900 Touch Control

■■■■■■■■

393

Remote
For automatic calibrations with Sample Processors, which are connec-
ted via the Remote Box.
USB
For automatic calibrations with a USB Sample Processor (see Chapter
32.6, page 492)
.

Standards
The concentrations of the standards as well as their units are defined
under [Standards].

Unit

Unit of the standard solutions used. For concentration measurements
without standard addition, this unit will be used.

Selection

mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L

Default value

ppm

Concentration standard 1

Concentration of the standard solution used.

Input range

00.001 - 999999.999

Default value

1,000

Concentration standard 2

Concentration of the standard solution used.

Input range

00.001 - 999999.999

Selection

off

Default value

off

Concentration standard 3

See Concentration standard 2.

Concentration standard 4

See Concentration standard 2.

Concentration standard 5

See Concentration standard 2.

Stirrer control
The parameters for the stirrer control during the calibration are edited
under [Stirrer control]. Ensure that a stirrer has been selected in the Edit
command / Stirrer
dialog.