29.15 Dosing and Liquid Handling

■■■■■■■■■■■■■■■■■■■■■■

460

■■■■■■■■

900 Touch Control

of two dosing devices so that the second dosing device is dosing while
the first one is being filled and vice-versa.

Dosing device

Selection of the MSB connector the dosing device is connected to. Every
MSB connector will always be displayed.

Selection

1 | 2 | 3 | 4 | off

Default value

off

Titrant

Selection of the titrant from the titrant list. We recommend always select-
ing the titrant. Titrants are defined under System

Titrants. You can

also enter a name which is not contained in the titrant list. When a deter-
mination is started there is a check whether the solution is contained in
list.

For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. For
the selected titrant, only the time interval for the GLP test of the buret unit
is checked at the start of the determination.

NOTE

The concentration, validity of the titer and the working life of the titrant
are ignored.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place.

Filling rate

Rate at which the dosing cylinder is filled.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488)
.

However, in order to ensure uninterrupted dosing, the following points
must be taken into account: