28.7 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

266

■■■■■■■■

900 Touch Control

Entering the bit pattern:

0 = line inactive

1 = line active

* = retain line status

p = set pulse

The output lines are always numbered from right to left, i.e. with the sig-
nal 00000000000001 line 0 is set. With a pulse, the length is set to 200
ms. If you wish to set pulses with other lengths, you have to define a cor-
responding template.

Entry

Bit pattern containing exactly 14 characters
or a max. of 24 characters for the name of
the template

Default value

00000000000001

Selection

Selection of the templates defined

[Set]

Set a defined output signal.

28.7

USB Sample Processor

Manual control

Sample Processor

With the function [Sample Processor], you can control a connected
Sample Processor manually. The following functions are available:

Moving a sample rack

Moving the lift

Defining specific lift positions (work position, rinse position, etc.)

Switching pumps on and off

Initializing the rack

Defining external positions outside of the sample rack (only with Swing
Head)