9.2 Reagent monitoring

■■■■■■■■■■■■■■■■■■■■■■

64

■■■■■■■■

900 Touch Control

Selection

off

Default value

off

Drift

This parameter is only visible with coulometric reagents.

During conditioning the measured drift has to be in the following range
for 2 min: 'specified drift value + 50 µg/min'.

Input range

1 - 999 µg/min

Selection

off

Default value

off

[Status]

Display the status overview of the current values of the reagent monitor-
ing.

[Reagent replacement]

Edit the parameters for the reagent exchange.

Dialog "Reagent monitoring / Status"
The current reagent monitoring values are displayed in this dialog.

[Reset]

Reset the values to zero.

Dialog "Reagent monitoring / Reagent replacement"
The parameters for the reagent exchange are defined in this dialog.

Reagent replacement

The reagent can either be exchanged manually or automatically.

Selection

manual | auto

Default value

manual

manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.
auto
If a monitored parameter has reached the limit set, the method defined
below is started automatically.

Memory

This parameter can only be edited with Reagent replacement = auto.

Memory location the method is loaded from. All memory locations are
selectable, even if they are currently not accessible.