9.2 Reagent monitoring
■■■■■■■■■■■■■■■■■■■■■■
64
■■■■■■■■
900 Touch Control
Selection
off
Default value
off
Drift
This parameter is only visible with coulometric reagents.
During conditioning the measured drift has to be in the following range
for 2 min: 'specified drift value + 50 µg/min'.
Input range
1 - 999 µg/min
Selection
off
Default value
off
[Status]
Display the status overview of the current values of the reagent monitor-
ing.
[Reagent replacement]
Edit the parameters for the reagent exchange.
Dialog "Reagent monitoring / Status"
The current reagent monitoring values are displayed in this dialog.
[Reset]
Reset the values to zero.
Dialog "Reagent monitoring / Reagent replacement"
The parameters for the reagent exchange are defined in this dialog.
Reagent replacement
The reagent can either be exchanged manually or automatically.
Selection
manual | auto
Default value
manual
manual
If a monitored parameter has reached the limit set, a message is being
displayed. Then the reagent has to be exchanged manually.
auto
If a monitored parameter has reached the limit set, the method defined
below is started automatically.
Memory
This parameter can only be edited with Reagent replacement = auto.
Memory location the method is loaded from. All memory locations are
selectable, even if they are currently not accessible.