7 stat titrations (stat), 1 start conditions, 7 stat titrations (stat) 29.7.1 – Metrohm 900 Touch Control Manual User Manual

Page 368: Start conditions

background image

29.7 STAT titrations (STAT)

■■■■■■■■■■■■■■■■■■■■■■

352

■■■■■■■■

900 Touch Control

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 32.3, page 489. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.

29.7

STAT titrations (STAT)

29.7.1

Start conditions

The parameters that are carried out before the start of titration are
defined under [Start conditions].

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is
added.

Input range

0 - 999999 s

Default value

0 s

Start volume

Volume that is dosed prior to the start of the titration.

Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum