9 measurements (meas), 1 measuring parameters, 9 measurements (meas) 29.9.1 – Metrohm 900 Touch Control Manual User Manual

Page 393: Measuring parameters

background image

■■■■■■■■■■■■■■■■■■■■■■

29 Parameters

900 Touch Control

■■■■■■■■

377

Switch off automatically

on | off (Default value: on)

If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.

29.9

Measurements (MEAS)

29.9.1

Measuring parameters

Under [Measuring parameters], the parameters influencing the
sequence of the entire measurement are defined.

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.

NOTE

A constant measured value is often only reached after a certain time, as
mixing and the reaction itself require a certain time. The response time
of an electrode can also increase with time, i.e. reaching a constant
measured value takes longer and longer. Drift-controlled measured
value acceptance is particularly advisable in such cases, as the measured
values are only accepted when equilibrium has almost been reached.

Measuring mode pH, U, Ipol and Conc:
Input range

0.1 - 999.0 mV/min

Default value

10.0 mV/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Measuring mode Upol:
Input range

0.01 - 99.90 µA/min

Default value

10.00 µA/min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.