29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
354
■■■■■■■■
900 Touch Control
29.7.2
Control parameters
The parameters for the control point are defined under [Control param-
eters].
Control point at
Measured value for the control point.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Measuring mode pH:
Input range
–20.000 - 20.000
Selection
off
Default value
off
Measuring mode U:
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection
50 µL/min | 100 µL/min | 500 µL/min | user
Default value
100 µL/min
50 µL/min
For reactions with an expected reaction rate in the range of 50 µL/min.
100 µL/min
For reactions with an expected reaction rate in the range of
100 µL/min.
500 µL/min
For reactions with an expected reaction rate in the range of
500 µL/min.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 12, page
356.
User-defined parameters
These parameters are only accessible when Titration rate = user.