beautypg.com

Metrohm 900 Touch Control Manual User Manual

Page 231

background image

■■■■■■■■■■■■■■■■■■■■■■

24 Results and more determination data

900 Touch Control

■■■■■■■■

215

[Measuring point list]

Display measuring point list of titrations and measurements (see "Measur-
ing point list", page 215)
.

Measuring point list
Results dialog: More data

View data Measuring point list

Results dialog: Curve

Display curve Measuring point list

The individual measured values are listed in the View data / Measuring
point list
dialog. You can configure which columns are displayed (button
[View]). These settings are saved separately for each mode (DET, MET,
etc.).

Column 1 / Column 2 / Column 3

Quantity displayed in the columns 1, 2 and 3. Both the selection of the
quantities and the default setting depend on the mode.

Selection

Volume | Measured value | Water | Bromine |
ERC | Delta meas.value | Time | Temperature |
Drift | Volume drift | Signal drift | Index | Mon-
itoring

Endpoint details
Results dialog: More data

View data Endpoint details

Results dialog: Curve

Display curve Endpoint details

Detailed information concerning each equivalence point or endpoint
determined is displayed in the View data / Endpoint details dialog.

The volume, the measured value, the equivalence point criterion ERC (only
DET, MET), the time and the temperature are specified for each equiva-
lence point or endpoint determined.