29.5 Coulometric Karl Fischer titrations (KFC)

■■■■■■■■■■■■■■■■■■■■■■

340

■■■■■■■■

900 Touch Control

NOTE

This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working
medium.

29.5.5

Cell

The parameters for the titration cell are defined under [Cell].

Generator electrode

Type of generator electrode.

Selection

without diaphragm | with diaphragm

Default value

without diaphragm

without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection

100 mA | 200 mA | 400 mA | auto

400 mA
Default value, when Generator electrode = without diaphragm.
auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint. Default value, when Generator
electrode
= with diaphragm.

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System

Reagents.

Selection

Selection of configured reagents | off

Default value

off

off
The reagent monitoring is not active.