beautypg.com

Metrohm 900 Touch Control Manual User Manual

Page 119

background image

■■■■■■■■■■■■■■■■■■■■■■

11 Device manager

900 Touch Control

■■■■■■■■

103

Rack offset

The rack offset is a production-related tolerance value between the upper
and lower sections of the rack. This value is determined when the rack is
adjusted and is displayed here. It can be changed if necessary.

Input range

–10.00 - 10.00 °

Default value

0.00 °

[Adjust rack]

Adjust the rack (see Chapter 11.6.2, page 107).

[Lift pos. tower 1]/[Lift pos. tower 2]

Define rack specific lift positions (see "Lift positions Tower 1/2", page
103)
.

[Special beakers]

Define rack positions as special beakers (see "Editing special beakers",
page 105)
.

Lift positions Tower 1/2
Sample rack list: Rack

Edit Lift pos. tower 1 / Lift pos. tower 2

You can define rack-specific lift positions in the dialog Edit rack data /
Lift positions tower 1/2
. These then apply for all rack positions except
those that have been defined as special beakers.

These positions can be moved to directly under manual control and with
the LIFT command. Only lift positions within the maximum stroke path can
be entered. This is defined in the device properties of the tower.