29 Parameters

900 Touch Control

■■■■■■■■

447

NOTE

This formula applies only for using a single dosing device. The dosing
rate is equal to the flow rate in the case of tandem dosing.

If the required dosing rate exceeds the maximum dosing rate for the
selected exchange unit or dosing unit (see Chapter 32, page 488), then a
larger cylinder volume must be selected.

Dosing criterion "Dosing rate/Dosing time"

Single dosing

Tandem dosing

V/mL

t/s

Dosing rate

Filling time

t

1

t

2

t/s

V/mL

Dosing rate

t

1

2

+ t

Figure 24

Dosing criterion "Dosing rate/Dosing time"

Volume

Volume which is dosed.

Input range

0.00000 - 99999.9 mL

Default value

10.0000 mL

Dosing rate

Rate at which it is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488)
.

Dosing time

Time while dosing is carried out, i.e. the filling time is not taken into
account.

Input range

0 - 999999 s

Default value

100 s