28.7 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

270

■■■■■■■■

900 Touch Control

External positions (only with robotic arm and mounted Swing Head,
see Chapter 28.7.3, page 272)
For each of the four possible external positions, a specific working
position can be defined. Shift and rinse positions can only be defined
for all four external positions together. A special position is not possi-
ble.

Proceed as follows to assign the current lift position to a specific lift posi-
tion:

1 Move to lift position

Enter the desired lift position in mm and tap on [Start].

The lift moves to the desired lift position.

2 Assign the lift position

Tap on [Assign lift pos.].
The dialog Lift / Assign lift position is displayed.

Select the desired specific lift position and tap on [Assign].

The current lift position is assigned to the specific lift position.

28.7.2

Moving to a rack position

Manual control: Sample Processor

Rack position