15.4 Output lines
■■■■■■■■■■■■■■■■■■■■■■
156
■■■■■■■■
900 Touch Control
Signal name
Output signal
Function
Meter Mode T
*********0010*
switches the 780/781 pH/Ion meter to tempera-
ture measurement and starts the measurement.
Meter Mode U
*********0011*
switches the 780/781 pH/Ion meter to voltage
measurement and starts the measurement.
Start device*
********p****p
starts Devices 1 and 2 (e.g. Titrino, Titrando, etc.)
*)
.
The signal is transmitted as a short pulse of 200
ms.
Start device1
*************p
starts Device 1 (e.g. Titrino, Titrando, etc.)
*)
.
The signal is transmitted as a short pulse of 200
ms.
Start device2
********p*****
starts Device 2 (e.g. Titrino, Titrando, etc.)
*)
.
The signal is transmitted as a short pulse of 200
ms.
Start Dos*
*****p*p******
starts Dosimat at Devices 1 and 2 (Titrino via "acti-
vate").
The signal is transmitted as a short pulse of 200
ms.
Start Dos1
*******p******
starts Dosimat at Device 1 (Titrino via "activate").
The signal is transmitted as a short pulse of 200
ms.
Start Dos2
*****p********
starts Dosimat at Device 2 (Titrino via "activate").
The signal is transmitted as a short pulse of 200
ms.
*)
In the case of the 780/781 pH/Ion meter, a result report is triggered.
[New]
Add a new output signal to the list, see following chapter.
[Delete]
Delete the selected output signal from the list.
[Edit]
Edit the data of the selected output signal, see following chapter.