32 Appendix

900 Touch Control

■■■■■■■■

511

Connect the curve simulator to an MSB connector on the
Titrando.

2 Configuring the Touch Control

Switch on the Touch Control.

In the System / Diagnosis dialog, tap on the [822 Curve Simu-
lator]
button.
The Diagnosis / 822 Curve Simulator dialog is displayed.

Activate the parameter Send dosing signals to MSB2 and con-
firm with [Set].

Tap on [

].

The signals from the dosing device will now be copied to the MSB
connector 2.

3 Simulating a titration

Create a titration method or load an existing one.

Tap on [

].

The titration is started. After the completion of the titration, the
results dialog is displayed.

4 Finishing the simulation mode

Switch off the Touch Control.

32.9.9

Service

System

Diagnosis Service

Access to the service functions is password-protected and only accessible
to Metrohm service technicians. We recommend that you have mainte-
nance performed on the system regularly by the Metrohm Service Depart-
ment. You can automatically monitor the time interval after which the
next service visit is due (see Chapter 13.5.1, page 137).