32 Appendix
900 Touch Control
■■■■■■■■
511
■
Connect the curve simulator to an MSB connector on the
Titrando.
2 Configuring the Touch Control
■
Switch on the Touch Control.
■
In the System / Diagnosis dialog, tap on the [822 Curve Simu-
lator] button.
The Diagnosis / 822 Curve Simulator dialog is displayed.
■
Activate the parameter Send dosing signals to MSB2 and con-
firm with [Set].
■
Tap on [
].
The signals from the dosing device will now be copied to the MSB
connector 2.
3 Simulating a titration
■
Create a titration method or load an existing one.
■
Tap on [
].
The titration is started. After the completion of the titration, the
results dialog is displayed.
4 Finishing the simulation mode
■
Switch off the Touch Control.
32.9.9
Service
System
▶ Diagnosis ▶ Service
Access to the service functions is password-protected and only accessible
to Metrohm service technicians. We recommend that you have mainte-
nance performed on the system regularly by the Metrohm Service Depart-
ment. You can automatically monitor the time interval after which the
next service visit is due (see Chapter 13.5.1, page 137).