29 Parameters
900 Touch Control
■■■■■■■■
301
Measuring mode pH:
Input range
0.10 - 9.99
Default value
0.50
Measuring mode U, Ipol:
Input range
1 - 999 mV
Default value
30 mV
Measuring mode Upol:
Input range
0.1 - 99.9 µA
Default value
2.0 µA
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
Selection
all | greatest | last | ascending | descending |
off
Default value
all
all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
No evaluation takes place.
Set windows
This parameter allows to recognize equivalence points only in a particular
range (so-called window) of the curve. Equivalence points outside this
window will not be recognized. A maximum of 9 windows can be
defined. Setting windows is advisable when interference and unnecessary
equivalence points are to be suppressed.
Selection
Measured value | Volume | off
Default value
off