29 Parameters

900 Touch Control

■■■■■■■■

301

Measuring mode pH:
Input range

0.10 - 9.99

Default value

0.50

Measuring mode U, Ipol:
Input range

1 - 999 mV

Default value

30 mV

Measuring mode Upol:
Input range

0.1 - 99.9 µA

Default value

2.0 µA

EP recognition

This parameter allows you to filter out only the equivalence points that are
being sought.

Selection

all | greatest | last | ascending | descending |
off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
No evaluation takes place.

Set windows

This parameter allows to recognize equivalence points only in a particular
range (so-called window) of the curve. Equivalence points outside this
window will not be recognized. A maximum of 9 windows can be
defined. Setting windows is advisable when interference and unnecessary
equivalence points are to be suppressed.

Selection

Measured value | Volume | off

Default value

off