Metrohm 900 Touch Control Manual User Manual
Page 341
■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
325
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).
Min. increment
Minimum volume increment that is dosed at the beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum increment, the slower the titration.
Input range
0.10 - 99.90 µL
Selection
minimum
Default value
minimum
Table 9
Default values of the predefined titration rates for KFT
Titration rate
slow
optimal
fast
Dynamics
– Ipol
– Upol
300.0 mV
40.00 µA
100.0 mV
10.00 µA
30.0 mV
5.00 µA
Max. rate
1.00 mL/min
maximum
maximum
Min. increment
minimum (= cylinder
volume/10,000)
minimum (= cylinder
volume/10,000)
5.00 µL
Stop criterion
Stop criterion
The titration is stopped when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be stopped. The stop conditions (see Chapter 29.4.4,
page 327) always lead to a stop, even if the stop criterion has not been
reached.
NOTE
In older instructions the delay time was usually defined as the stop cri-
terion. The same delay time may however result in different stopping
time points – because of different smallest possible volume increments
(depending on the cylinder volume). In contrast, if the stop drift is used,
the titration will always be stopped at the same curve slope dV/dt.
Selection
drift | time | rel. drift | off
Default value
drift