29 Parameters

900 Touch Control

■■■■■■■■

395

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value

0 s

Max. waiting time

If the signal drift has been switched off or has not yet been reached, then
the measured value will be accepted when the maximum waiting time has
elapsed.

Input range

0 - 999999 s

Default value

10 s

Selection

off

off
The measurement is continued endlessly.

Measuring temp.

Temperature entered manually. The temperature is being continuously
measured when a temperature sensor is connected and when Temp.
measurement
= automatic or continuous is defined (see sensor dialog
of the command).

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Standard
The characteristics of the calibration standard used are defined under
[Standard]. These values will be suggested in the request after the start
of the calibration.

Request at calibration start

on | off (Default value: off)

When this parameter is activated, then entering the following parameters
will be requested after the start.

Conductivity standard

Conductivity of the calibration standard used.

Input range

0.00000 - 2000.00 mS/cm

Default value

12.8800 mS/cm