beautypg.com

29 parameters, 1 dynamic equivalence point titrations (det), 1 start conditions – Metrohm 900 Touch Control Manual User Manual

Page 292: Start conditions

background image

29.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

276

■■■■■■■■

900 Touch Control

29 Parameters

NOTE

For most of the numerical parameters, the result of a calculation can
also be utilized in place of a number. A description of this can be found
in the appendix (see Chapter 32.5, page 491).

29.1

Dynamic equivalence point titrations (DET)

29.1.1

Start conditions

The parameters that are carried out before the start of titration are
defined under [Start conditions]. The start conditions are processed in
the following sequence:

1.

Start volume

2.

Start measured value

3.

Start slope

4.

Pause

Start volume

Volume that is dosed prior to the start of the titration.

Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Dosing rate

Rate at which the start volume is dosed.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488)
.

Pause

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume. The time entered here is
only running when all start conditions have been fulfilled.

Input range

0 - 999999 s

Default value

0 s