beautypg.com

5 conditioning a titration cell – Metrohm 900 Touch Control Manual User Manual

Page 280

background image

28.5 Conditioning a titration cell

■■■■■■■■■■■■■■■■■■■■■■

264

■■■■■■■■

900 Touch Control

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

NOTE

If a stirrer is used in the determination run, it can nevertheless be con-
trolled manually at the same time. For example, you can use the man-
ual control to reduce the stirring rate of a stirrer which has been started
in the determination run.

28.5

Conditioning a titration cell

Manual control

Conditioning

In manual control, you have the option of permanently conditioning one
titration cell for Karl Fischer titrations while carrying out a Karl Fischer
titration in a second titration cell.

Proceed as follows:

1 Create a method

Create a Karl Fischer method.

Save the method.

2 Load the method

Select the memory location where the desired method is saved.

Tap on the input field File name.
The method table with the saved methods is opened.

Select the desired method.

3 Start the conditioning

Tap on [Start].

The working medium is titrated to the endpoint. The status is kept
stable.