28.5 Conditioning a titration cell
■■■■■■■■■■■■■■■■■■■■■■
264
■■■■■■■■
900 Touch Control
The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:
■
"+": counterclockwise rotation
■
"–": clockwise rotation
Input range
–15 - 15
Default value
8
NOTE
If a stirrer is used in the determination run, it can nevertheless be con-
trolled manually at the same time. For example, you can use the man-
ual control to reduce the stirring rate of a stirrer which has been started
in the determination run.
28.5
Conditioning a titration cell
Manual control
▶ Conditioning
In manual control, you have the option of permanently conditioning one
titration cell for Karl Fischer titrations while carrying out a Karl Fischer
titration in a second titration cell.
Proceed as follows:
1 Create a method
■
Create a Karl Fischer method.
■
Save the method.
2 Load the method
■
Select the memory location where the desired method is saved.
■
Tap on the input field File name.
The method table with the saved methods is opened.
■
Select the desired method.
3 Start the conditioning
■
Tap on [Start].
The working medium is titrated to the endpoint. The status is kept
stable.